An intelligent grasper to provide real-time force feedback to shorten the learning curve in laparoscopic training

Abstract

Background

A lack of force feedback in laparoscopic surgery often leads to a steep learning curve to the novices and traditional training system equipped with force feedback need a high educational cost. This study aimed to use a laparoscopic grasper providing force feedback in laparoscopic training which can assist in controlling of gripping forces and improve the learning processing of the novices.

Methods

Firstly, we conducted a pre-experiment to verify the role of force feedback in gripping operations and establish the safe gripping force threshold for the tasks. Following this, we proceeded with a four-week training program. Unlike the novices without feedback (Group A2), the novices receiving feedback (Group B2) underwent training that included force feedback. Finally, we completed a follow-up period without providing force feedback to assess the training effect under different conditions. Real-time force parameters were recorded and compared.

Results

In the pre-experiment, we set the gripping force threshold for the tasks based on the experienced surgeons’ performance. This is reasonable as the experienced surgeons have obtained adequate skill of handling grasper. The thresholds for task 1, 2, and 3 were set as 0.731 N, 1.203 N and 0.938 N, respectively. With force feedback, the gripping force applied by the novices with feedback (Group B1) was lower than that of the novices without feedback (Group A1) (p < 0.005). During the training period, the Group B2 takes 6 trails to achieve gripping force of 0.635 N, which is lower than the threshold line, whereas the Group A2 needs 11 trails, meaning that the learning curve of Group B2 was significantly shorter than that of Group A2. Additionally, during the follow-up period, there was no significant decline in force learning, and Group B2 demonstrated better control of gripping operations. The training with force feedback received positive evaluations.

Conclusion

Our study shows that using a grasper providing force feedback in laparoscopic training can help to control the gripping force and shorten the learning curve. It is anticipated that the laparoscopic grasper equipped with FBG sensor is promising to provide force feedback during laparoscopic training, which ultimately shows great potential in laparoscopic surgery.

Peer Review reports

Introduction

In the past few decades, Minimally Invasive Surgery (MIS) has changed the surgical process [1,2,3], greatly improving the implementation of MIS in general surgery, gynecology, cardiothoracic surgery, colorectal surgery and urological surgery [4,5,6,7]. MIS has many advantages, such as small wound, quickly recovery, etc [1, 8]. However, the lack of force feedback, even completely lose in Robot-assisted minimally invasive surgery (RMIS), during operation restricts its development [9] and affects the training of novices’ operating during MIS to some extent. The lack of force feedback in MIS leads novices to undergo more training before operating successfully on patients [2]. Generally, the skill acquisition involves a steep learning curve [10]. Most residents who receive training in laparoscopic surgical skills take more than 3 years to master the operation skills [11]. Performing complex surgical tasks in laparoscopic surgery requires more precise control and extensive training [12].

Indeed, in order to meet the precise requirements of MIS, more and more simulation training and education methods outside of the operation room are proposed to improve the surgical skills of surgeons, particularly novice surgeons. With the increasing availability and use of laparoscopic training models, more and more novice surgeons are able to acquire the necessary skills for MIS through simulation-based training programs. This has the potential to improve patient outcome by reducing the risk of surgical errors and complications during actual procedures. There have been a series of training models developed for simulation-based surgical training. Currently, the most widely used simulation methods can be classified into three categories: box training (BT), virtual reality (VR) and augmented reality (AR) training. However, traditional BT models do not provide force feedback, which limits their ability to provide a realistic training context and objective results. On the other hand, VR training is able to provide objective results, but these results are typically not available to operators in real-time. AR that provides both haptic feedback and objective results during training [13] is costly. Therefore, it is essential to develop cost-effective training systems that can provide force feedback in order to enhance the effectiveness and realism of surgical simulation training. There are some systems appearing, which however rely on external force measuring platform and may lead to imprecise results [14]. For example, Luis et al. [15] used an intelligent trainer equipped with a gripping sensor to measure the gripping force, but the measuring force is not really accurate. Hardon et al. [16] use a box trainer with a built-in force tracking system to monitor the force to assess the operation skill of residents. In fact, due to the status of the usage of the training system, the training in MIS offering to novices is limited [17]. Regarding the current laparoscopic training system with force feedback, its high cost of training and absence of objective real-time assessments have resulted in a prolonged learning curve for novice surgeons [18]. Therefore, there is an urgent need to develop an effective and costless training system. In the training processing of MIS, it is preferable to use a laparoscopic grasper with real-time force feedback to achieve targeted training and increase the efficiency of every surgical operation. The training process will be effectively standardized and the training period will be shortened [19].

In the past few years, fiber Bragg grating (FBG) sensors have been widely applied in minimally invasive surgery to offer force feedback because of their small size, high sensitivity, good biocompatibility, light weight, immunity to electromagnetic interference (EMI), etc [20,21,22,23,24]. The surgical instrument integrated with FBG sensor provides the surgeon with force information during MIS, facilitating more precise and accurate operations. For example, Li et al. [25] proposed a three-axis tactile probe based on fiber grating, which can accurately identify long blood vessels in the prosthesis and locate the wrapped tissue in three dimensions. Furthermore, they verified its effectiveness and feasibility in isolated porcine kidney tissue. Xue et al. [26] introduced FBGs to grooves in the laparoscopic surgery robot, which is used to estimate gripping force and perform precise force control. Besides, Imbrie-moore et al. [27] mounted a pig mitral valve in a cardiac simulator, each valve was repaired with Teflon sutures. In their work, an FBG sensor was used to measure real-time suture force. In addition, Scott et al. [28] developed an FBG-based sensor and measured the force at the tip of the electrode array during insertion into the cochlea in real time in guinea pigs. Although FBG has been proposed for use in medical surgery, the additional value of force feedback based on FBG in MIS training has not been established.

In our previous study, we designed an intelligent laparoscopic grasper integrated with an FBG-based tactile sensor, which can provide real-time force feedback to the novice operators and has shown excellent performance in the laparoscopic training box [29]. As a continuation of this work, we utilized the laparoscopic grasper to provide real-time force feedback to the trainers during laparoscopic training, allowing for quantification of force information obtained during training. Results indicate that training system with an FBG force sensor has significant potential to shorten the learning curve of laparoscopic training by providing real-time force feedback to trainees through an intelligent laparoscopic grasper.

Methods

The procedures, methods, and consent forms employed in this study received approval from the Ethics Review Board of Zhujiang Hospital, Southern Medical University. The training program, which was based on basic laparoscopic skills (FLS) closely simulated real-world clinical circumstance.

The proposed laparoscopic training system and gripping tasks setting

The proposed training system is shown in Fig. 1. To imitate the laparoscopic surgery, we use a validated Lap Game box trainer (Lap Game Inc., Hangzhou, China). The device (Fig. 1) includes a light source, an internal camera for imaging and a display screen for operation on the computer. Different from traditional laparoscopic training box, we used a previously designed smart laparoscopic grasper with fiber Bragg grating sensor during training [29]. The intelligent laparoscopic grasper is used to clamp the specific objects and the real-time force information obtained by the fiber Bragg grating sensor was demodulated by an optical spectrometer (I-MON 51 USB, Ibsen Photonics, Denmark). In principle, FBG can reflect a specific wavelength (also called Bragg wavelength λ B) of broadband light , which is determined by the effective index of refraction (neff) of the fundamental mode propagating in the fiber core and the grating period (Λ), as expressed in Eq. (1). When there is a certain force applied to the FBG, the Bragg wavelength would show a corresponding shift due to the deformation of the grating as well as a change in the refractive index. This relationship can be defined by Eq. (2), in which Pe is photo-elastic coefficient, η is a factor of force transferred to strain and F is the applied force.

$${\uplambda}_B=2{n}_{\textrm{eff}}\Lambda$$
(1)
$$\Delta {\uplambda}_B=2{n}_{\textrm{eff}}\Lambda \left(1+{P}_e\right)\cdot \left(\upeta F\right)$$
(2)
Fig. 1
figure 1

Instrument connection diagram

In our study, we can obtain the gripping force by demodulating the Bragg wavelength change in the reflection spectrum detected by the spectrometer during the operation process. The real-time force information is displayed and stored on the computer.

The experiment consisted of 3 gripping transfer tasks (Table 1). These tasks are designed based on basic laparoscopic skills (FLS) training [30] and close to real clinical circumstance [31].

Table 1 Gripping tasks setting

Participant selection and baseline characteristics

We recruited medical students with no prior training in laparoscopic surgery from our institution’s medical school through a virtual announcement. Gynecologists with extensive experience (having perform over 100 advanced procedures) were selected as surgeons with experience for the study. During the study, the participants included 6 experienced surgeons and 42 novices. Novices were divided randomly into novices without force feedback (Group A) and novices with force feedback (Group B). All participants filled out the personal questionnaire before the experiment. The basic information of all subjects was documented and recorded in Table 2. In order to ensure that the novice participants had a similar level of laparoscopic surgery skills, 42 novices were required to complete Task 1 without feedback for baseline assessment. Novices at the same level were included in this study.

Table 2 Personal information of the participants

Study protocol

Preliminary experiment

We conducted a pre-experiment to verify the role of force feedback in laparoscopic gripping and set the threshold for the gripping tasks. Without providing force feedback, the experienced surgeons (Group C) and six novices of Group A (Group A1) were asked to complete three tasks to see whether there is some difference between the two and further determine the threshold of the tasks. The threshold of the task was defined as the force level of safe grasping. We set the average of the maximum gripping force of the experienced surgeons as the threshold of the tasks. The thresholds for task 1, task 2 and task 3 are 0.731 N, 1.203 N and 0.938 N respectively. Then, with real-time force feedback, six novices of Group B (Group B1) carried out three tasks. Different from novices without force feedback (Group A1), novices with force feedback (Group B1) conducted the tasks with real-time force feedback, meaning that novices with force feedback can quantify the grasping force during the tasks and adjust the force they used according to the threshold of the tasks. Throughout the operation, all the grasping data was collected by the computer. A comparison of the grasping force between Group A1 and Group B1 was conducted. The purpose of the preliminary experiment is to determine whether force feedback during gripping operations can have a positive influence on MIS. Finally, laparoscopic training sessions were conducted for the remaining novices in the laparoscopic training box.

Training program

The other thirty novices who are at the same level in laparoscopic surgery entered the laparoscopic training. They are Group A2 and Group B2, i.e., the other 15 novices of Group A and Group B, respectively. A four-week laparoscopic training was conducted. Taking schedule of the novices into consideration, we ensured that all novices completed an equal number of gripping trials during this training period. In our study, 30 novices were assigned to conduct 10 trials at different time intervals. Each trial required them to successfully complete the task three times. It is important to note that all novices utilized the same smart laparoscopic grasper. As for the Group B2, with the real-time force feedback, participants were able to quantitatively measure the force applied during tasks and adjust it in time according to the predefined threshold. In contrast, Group A2 completed the training without force feedback, relying solely on subjective perception for force adjustment. All the grasping force data during the whole training was collected by the FBG force sensor.

Follow-up test

Follow-up testing was carried out 1 week after the completion of training to evaluate the retention of grasping skills acquired during the training period. Both Group A2 and Group B2 were require to complete the task without force feedback in the follow-up period. Besides, all participants were asked to complete the NASA Task Load Index [32]. Only Group B2 filled out the Force Feedback System Evaluation Survey additionally. Figure 2 shows the schematic flowchart of the study protocol.

Fig. 2
figure 2

Schematic flowchart of the study protocol

Outcome evaluation

The maximum absolute force and the standard deviation of the gripping force collected during the experiment are utilized to evaluate the effectiveness of the training. Additionally, scales filled out by the participants are analyzed at the same time to assess the training system from subjective point of view. A detailed overview and description of these parameters are provided in Table 3. In addition, in order to better analyze the control and learning progress of two groups during the training period, we plotted individual learning curves for each group and conducted a comparison analysis. Specifically, we recorded the maximum and standard deviation of the gripping force during 10 laparoscopic training sessions of novices under different feedback conditions. Additionally, we included the final follow-up stage where both groups performed gripping operations without force feedback. The training curves were plotted at each data point. Drawing the learning curves during the laparoscopic training allows for a more intuitive observation of the impact of real-time force feedback. All participants were asked to fill out a questionnaire and NASA Task Load Index after completion of the four-week training to obtain information about their general impression of the force feedback system and training tasks. General comments on the training were obtained from participants and presented using a 5 points Likert scale.

Table 3 Description of outcome evaluation

Data were analyzed using IBM SPSS statistical version 23.0 (IBM Corp, Armonk, NY). The t-test was employed for normally distributed data, while nonparametric Mann-Whitney U test was used for non-normally distributed data. Basically, a probability of p<0.05 was considered statistically significant [33].

Results

Baseline assessment of the novices

At the baseline evaluation, all novices had no prior experience in laparoscopic operation. Figure 3 shows the box plots of gripping force that represents the baseline assessment of the novices (n = 42). Obviously, none of the peak, mean and standard deviation of the gripping force showed a significant difference among the novices, indicating that the two groups of novices were at the same level of laparoscopic grasping (p = 0.653, 0.996, 0.831 respectively).

Fig. 3
figure 3

Box plots of the gripping force among novices for baseline assessment

Analysis results of preliminary experimental

In the preliminary experiments, Table 4 presents the gripping force of three groups in different tasks. The statistical results comparing the three groups are also summarized. Taking task 1 as an example, the average of the maximum gripping force for experienced surgeons was 0.731 N, while novices without feedback recorded 3.686 N. The standard deviation of the gripping force for the two groups was 0.096 and 0.468 respectively. The force values for the task exhibited significantly differences (p < 0.001). The results indicated that the experienced surgeons exhibited significantly lower force and better stability of control than the novices without feedback. Similar results were observed across other tasks. Therefore, we established the average maximum value of the experienced surgeons as the threshold for the tasks. The threshold for task 1, task 2 and task 3 were determined to be 0.731 N, 1.203 N and 0.938 N respectively. With providing real-time force feedback to Group B1, the maximum and the standard deviation of the gripping force of task 1 is 0.979 N and 0.112, while is 3.686 N and 0.486 of Group A1. In the comparison between Group A1 and Group B1, the force values of all tasks were compared, resulting in a p value of less than 0.05. This indicates that the introduction of force feedback has obvious advantages in better maintaining gripping force. Additionally, the maximum value of all tasks in novices with feedback and experienced surgeons is significantly different(p<0.001). Although novices with feedback exhibit better control of gripping force, this result suggests that, compared to the experienced surgeons, novices still have room for improvement in controlling gripping force. Therefore, implementing a standardized training process is necessary to further enhance the control of gripping force.

Table 4 Statistical results between different groups in all the tasks

Analysis results of the training program

To gain a deeper understanding of the impact of force feedback on the learning curve, we presented the learning curves for the entire training process and subsequent follow-up trial. Figure 4 illustrates that in both groups, the maximum gripping force and standard deviation of the gripping force exhibit a gradual decrease throughout the training period.

Fig. 4
figure 4

(A) and (B) represent the learning curves of maximum force and standard deviation of the gripping force, respectively (SD, standard deviation)

In Fig. 4A, compared to Group A2, at the sixth training trial, the maximum gripping force of Group B2 is 0.635 N, which was below the threshold, while Group A2 still failed to reach the level below the threshold even by the tenth training. Furthermore, in the final training trial, the maximum gripping force of Group B2 was only 0.363 N, while the maximum gripping force of Group A2 is 0.765 N, still surpassing the threshold level. Obviously, Group B2 demonstrated a significantly shorter learning curve compared to Group A2. Therefore, the use of a smart grasper with an FBG force sensor can effectively expedite the training process and lead to a shorter curve.

In terms of the standard deviation of gripping force during training, which indicates gripping stability, Fig. 4B demonstrates that Group B2 exhibits a smaller standard deviation compared to Group A2. This implies that in the whole training process, Group B2 demonstrates superior stability and less fluctuation in gripping force than Group A2.

In the comparison of training effects in each stage of training stages (Table 5), the gripping force exhibited significant differences between Group A2 and Group B2 (p<0.05).These results indicate that novices who received feedback performed better during laparoscopic gripping training.

Table 5 Statistical results between different groups in training trials

Analysis results of follow-up test

Throughout the follow-up period, there was no significant decline in force learning, and the gripping force continued to decrease. Without providing force feedback, the maximum gripping force of Group A2 was 0.706 N, while Group B2 was only 0.316 N. Notably, the maximum gripping force in both groups fell below the threshold level, indicating a better retention of force. Group B2 maintained superior gripping performance compared to Group A2 after training. Additionally, the standard deviation of the gripping force for Group A2 increased slightly to 0.089 N, whereas Group B2’s standard deviation continued to decrease to 0.035 N.

In Fig. 5, we employed statistical methods to analyze the impact of force feedback on laparoscopic training across the baseline stage, training stage, and follow-up stage. During the baseline stage, the results indicate that there were no statistically significant differences in both the maximum and standard deviation of gripping force between Group A2 and Group B2(p>0.05). These findings suggest that novices in both groups performed at a similar level of gripping proficiency. Compared to the experienced surgeons, novices tend to apply higher gripping force during the initial stage of an operation. This often causes unnecessary tissue damage in clinical operation, so it is essential to train novices and verify whether incorporating force feedback in the operation has a positive effect on novices’ training. During the training phase, the results showed that the maximum force and standard deviation of the novices decreased during the training process. Moreover, the maximum gripping force of Group B2 on day 1, 5 and 10 were 1.281 N, 0.796 N, 0.363 N respectively while they were 1.811 N, 1.263 N, 0.765 N of Group A2. The difference between Group A2 and Group B2 was statistically significant (p < 0.05), indicating that the training effect was better with providing force feedback. Without providing force feedback, novices completed the follow-up period. The maximum gripping force of Group A2 was 0.706 N, while Group B2 was only 0.316 N. The difference between Group A2 and Group B2 was statistically significant (p < 0.05), indicating that Group B2 maintained superior gripping performance compared to Group A2 after training.

Fig. 5
figure 5

Gripping force at different stages under various conditions. (a) The maximum gripping force of Group C, Group A2 and Group B2 during baseline assessment stage, training stage and follow-up stage, (b) the standard deviation of gripping force of Group C, Group A2 and Group B2 during these three stages

All thirty novices enrolled in the training period completed the questionnaire. The evaluation scales for both the force feedback system and force feedback training utilized a 5 points Likert scale, which a maximum score of 5 points per question. As illustrated in Table 6, the system design received a score of 4.53 ± 0.52, indicating high satisfaction with the system. Similarly, regarding the visualization, instrumentation, user-friendliness, task description, force accuracy of feedback, and necessity of the system, scores were closed to the maximum of 5 points. Regarding for the evaluation of the force feedback training, participants self-rated an improvement in their technical skills and self-confidence as a result of the training. Moreover, trainees expressed that they found the training to be necessary and stated their willingness to recommend it to others. In term of the results from The NASA Task Load Index (Table 7), which assesses novices’ subjective feeling during the training, it was observed that novices who received feedback obtained lower scores in mental stress, psychological burden, operating time, and effort compared to those without feedback. These results suggest that the use of the smart laparoscopic grasper, which provides real-time force feedback, can lead to a more comfortable and confident gripping experience for novices. When it comes to the feeling of satisfaction, the scores of novices with feedback was higher than those without feedback, indicating that the incorporation of feedback resulted in a more positive emotional experience during the training.

Table 6 Survey outcomes
Table 7 The NASA task load index

Discussion

This study included a preliminary experiment to evaluate the impact of real-time force feedback on gripping, which demonstrated improved control of gripping force during MIS. Furthermore, a four-week laparoscopic training was conducted to confirm that force feedback helps to shorten the learning curve in laparoscopic training. Through the use of proposed smart laparoscopic grasper integrated with an FBG-based tactile sensor, novice participants were able to achieve real-time force feedback resulting in a shorter learning curve and an improved control of the gripping force during training. As a further assessment, a follow-up gripping operation was carried up, and several scales were performed to gauge the participants’ subjective perception of the training process. During the follow-up period, novices who had trained with force feedback demonstrated improved control of gripping even without the provision of force feedback. As for the result of scales assessment, both the training system itself and the emotional experience of the training process, as well as the benefits derived from the training, were highly rated. Accounting for the novices may vary in different gripping levels at the beginning, which somehow influence the result of the study, we did a baseline assessment and only participants at the same level were enrolled in the experiment. While individual abilities may differ, there were no significant differences in the overall evaluation. This suggests that the shortened learning curve is mainly due to the use of force feedback instead of the differences in individual participants’ abilities. Besides, taking the schedule of the novices into consideration, we ensured that all novices had the same number of gripping trails during the four-week training. Novices were allowed to take their time to complete the gripping tasks, accommodating their individual pace.

As pointed out by Hopper et al. [34], the ideal surgical learning curve (LC) tends to show a steep curve at first, which then fades out to a more gradual LC as the plateau phase approached. The technical skills are sufficiently established to operate independently and safely at last [35]. In our study, both the maximum force and the standard deviation of force were observed to have significantly decreased throughout the training period, as shown in Fig. 4. Furthermore, during the follow-up period, there was no significant increase in gripping force. On the contrary, gripping force further decreased due to improved control of the grasper. These results confirmed that the training effect can be preserved, and the LC of novices reach a fairly stable stage. Furthermore, compared to novices without feedback (Group A2), novices with feedback (Group B2) were observed to achieve better control of gripping force during the training. The learning curve of Group B2 was also noticeably shortened in comparison to Group A2. These results further demonstrate the effectiveness of real-time force feedback in improving the learning process and enhance the performance of novices in laparoscopic training. According to the results of our study, the gripping skills in the training box of the novices are improved because of the training and real-time force feedback. The integration of a smart laparoscopic grasper with force feedback capability contributed to maintaining optimal gripping force levels and results in a shorten the learning curve. These results highlight the potential benefits of incorporating force feedback technology in laparoscopic training to improve the performance and skill acquisition of novices. The introduction of force feedback is a feasible and valuable mechanism for enhancing laparoscopic training among novices. In the future, it should be considered as an integrated part of laparoscopic training programs, allowing for the development of individualized courses based on trainee’s learning curve. Eventually, the cost of laparoscopic learning training is expected to decrease. Besides, implementing a threshold for gripping tasks serves to raise novice’s awareness about the potential risks associated with applying excessive forces in the box trainer, which can lead to unnecessary tissue damage. Real-time force feedback allows them to adjust their gripping force in time according to the threshold, which is likely to be beneficial in clinical surgery [36]. If laparoscopy novices can hand and adjust their force in time, tissue damage [37] will be reduced. Furthermore, by analyzing the learning curve of the novices in the laparoscopic training, we can distinguish skill levels of surgeons to some extent [38]. Thus, a more targeted and individualized training plan can be established to realize precise training [39]. The promising results demonstrate the feasibility and value of integrating force feedback into laparoscopic training for novices. This study highlights the significant potential of the smart laparoscopic grasper with FBG force sensor in improving training, quality and reducing the learning curve. Ultimately, it offers a robust force feedback system for MIS.

In summary, the results of our study show that the utilization of an intelligent laparoscopic grasper with real-time force feedback in laparoscopic training contributes to improve the control of gripping force and a shortened learning curve.

A few limitations are encountered in the research. First of all, two screens may affect the mental and cause distraction in the process of gripping. The real-time force feedback of the training system needs to be further optimized by setting an operational screen in a form of “traffic lights” [40] or audio reminders. Secondly, clinical operations involve multiple gripping motion, and training tasks should be designed to imitate these movements. Finally, in-vivo experiments will be necessary to determine whether the positive effects observed in our study can be transferred to real-time scenarios. While our current results are not based on in-vivo testing, they provide promising evidence of the potential benefits of incorporating force feedback into laparoscopic training.

Conclusion

In conclusion, using a grasper providing real-time force feedback in laparoscopic training can help to control the gripping force and shorten the learning curve. It is anticipated that the laparoscopic grasper equipped with fiber Bragg grating sensor is promising to provide force feedback during laparoscopic training, which ultimately demonstrating significant potential in the field of laparoscopic surgery.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

MIS:

Minimally Invasive Surgery

RMIS:

Robot-assisted Minimally Invasive Surgery

BT:

Box Training

VR:

Virtual Reality

AR:

Augmented Reality

FBG:

Fiber Bragg grating

EMI:

Electromagnetic Interference

LC:

Learning Curve

References

  1. Othman W, Lai ZA, Abril C, Barajas-Gamboa JS, Corcelles R, Kroh M, et al. Tactile sensing for minimally invasive surgery: conventional methods and potential emerging tactile technologies. Front Robot AI. 2021;8:705662.

    Article  PubMed  Google Scholar 

  2. Jung Kim MM, Kim H, Srinivasan MA. Haptics in minimally invasive surgical simulation and training. IEEE Comput Graph Appl. 2004;24:56–64.

    Article  PubMed  Google Scholar 

  3. De Win G, Van Bruwaene S, Allen C, De Ridder D. Design and implementation of a proficiency-based, structured endoscopy course for medical students applying for a surgical specialty. Adv Med Educ Pract. 2013;4:103–15.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Zhu R, Marechal M, Suetsugu M, Yamamoto I, Lawn MJ, Matsumoto K, et al. Research and Development of testing device for evaluating force transmission and grasping pressure of laparoscopic forceps. Sensor Mater. 2019;31:4205–14.

    Article  Google Scholar 

  5. Ukai T, Tanaka Y, Fukuda T, Kajikawa T, Miura H, Terada Y. Softness sensing probe with multiple acoustic paths for laparoscopic surgery. Int J Comput Assist Radiol Surg. 2020;15:1537–47.

    Article  PubMed  Google Scholar 

  6. Ebina K, Abe T, Higuchi M, Furumido J, Iwahara N, Kon M, et al. Motion analysis for better understanding of psychomotor skills in laparoscopy: objective assessment-based simulation training using animal organs. Surg Endosc. 2021;35:4399–416.

    Article  PubMed  Google Scholar 

  7. Bandari N, Dargahi J, Packirisamy M. Image-based optical-Fiber force sensor for minimally invasive surgery with ex-vivo validation. J Electrochem Soc. 2020;167:127504.

    Article  CAS  Google Scholar 

  8. Othman W, Vandyck KE, Abril C, Barajas-Gamboa JS, Pantoja JP, Kroh M, et al. Stiffness assessment and lump detection in minimally invasive surgery using in-house developed smart laparoscopic forceps. IEEE J Transl Eng Health Med. 2022;10:2500410.

    Article  PubMed  Google Scholar 

  9. Smyk NJ, Weiss SM, Marshall PJ. Sensorimotor oscillations during a reciprocal touch paradigm with a human or robot partner. Front Psychol. 2018;9:2280.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Halim J, Jelley J, Zhang N, Ornstein M, Patel B. The effect of verbal feedback, video feedback, and self-assessment on laparoscopic intracorporeal suturing skills in novices: a randomized trial. Surg Endosc. 2021;35:3787–95.

    Article  PubMed  Google Scholar 

  11. Wenger L, Richardson C, Tsuda S. Retention of fundamentals of laparoscopic surgery (FLS) proficiency with a biannual mandatory training session. Surg Endosc. 2015;29:810–4.

    Article  PubMed  Google Scholar 

  12. Horeman T, van Delft F, Blikkendaal MD, Dankelman J, van den Dobbelsteen JJ, Jansen FW. Learning from visual force feedback in box trainers: tissue manipulation in laparoscopic surgery. Surg Endosc. 2014;28:1961–70.

    Article  PubMed  Google Scholar 

  13. Botden SM, Torab F, Buzink SN, Jakimowicz JJ. The importance of haptic feedback in laparoscopic suturing training and the additive value of virtual reality simulation. Surg Endosc. 2008;22:1214–22.

    Article  PubMed  Google Scholar 

  14. Hernandez R, Onar-Thomas A, Travascio F, Asfour S. Attainment and retention of force moderation following laparoscopic resection training with visual force feedback. Surg Endosc. 2017;31:4805–15.

    Article  PubMed  Google Scholar 

  15. Olivas-Alanis LH, Calzada-Briseno RA, Segura-Ibarra V, Vazquez EV, Diaz-Elizondo JA, Flores-Villalba E, et al. LAPKaans: tool-motion tracking and gripping force-sensing modular smart laparoscopic training system. Sensors (Basel). 2020;20(23):6937.

    Article  ADS  CAS  PubMed  Google Scholar 

  16. Hardon SF, Horeman T, Bonjer HJ, Meijerink W. Force-based learning curve tracking in fundamental laparoscopic skills training. Surg Endosc. 2018;32:3609–21.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Rekman JF, Alseidi A. Training for minimally invasive cancer surgery. Surg Oncol Clin N Am. 2019;28:11–30.

    Article  PubMed  Google Scholar 

  18. Elessawy M, Mabrouk M, Heilmann T, Weigel M, Zidan M, Abu-Sheasha G, et al. Evaluation of laparoscopy virtual reality training on the improvement of Trainees' surgical skills. Medicina (Kaunas). 2021;57:130.

    Article  PubMed  Google Scholar 

  19. Dai Y, Abiri A, Pensa J, Liu S, Paydar O, Sohn H, et al. Biaxial sensing suture breakage warning system for robotic surgery. Biomed Microdevices. 2019;21:10.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Deng Y, Yang T, Dai S, Song G. A miniature Triaxial Fiber optic force sensor for flexible Ureteroscopy. IEEE Trans Biomed Eng. 2021;68:2339–47.

    Article  ADS  PubMed  Google Scholar 

  21. Abushagur AA, Arsad N, Reaz MI, Bakar AA. Advances in bio-tactile sensors for minimally invasive surgery using the fibre Bragg grating force sensor technique: a survey. Sensors. 2014;14:6633–65.

    Article  ADS  CAS  PubMed  PubMed Central  Google Scholar 

  22. He X, Handa J, Gehlbach P, Taylor R, Iordachita I. A submillimetric 3-DOF force sensing instrument with integrated fiber Bragg grating for retinal microsurgery. IEEE Trans Biomed Eng. 2014;61:522–34.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Ping Z, Zhang T, Gong L, Zhang C, Zuo S. Miniature flexible instrument with fibre Bragg grating-based Triaxial force sensing for intraoperative gastric Endomicroscopy. Ann Biomed Eng. 2021;49:2323–36.

    Article  PubMed  Google Scholar 

  24. Bandari N, Dargahi J, Packirisamy M. Tactile sensors for minimally invasive surgery: a review of the state-of-the-art, applications, and perspectives. IEEE Access. 2020;8:7682–708.

    Article  Google Scholar 

  25. Li T, Pan A, Ren H. Reaction force mapping by 3-Axis tactile sensing with arbitrary angles for tissue hard-inclusion localization. IEEE Trans Biomed Eng. 2021;68:26–35.

    Article  PubMed  Google Scholar 

  26. Xue R, Ren B, Huang J, Yan Z, Du Z. Design and evaluation of FBG-based tension sensor in laparoscope surgical robots. Sensors (Basel). 2018;18:2067.

    Article  ADS  PubMed  Google Scholar 

  27. Aminebili-moore ZY, Yaoming P, Paulsen WH, Yao-jie W. Artificial papillary muscle device for off pump transapical mitral. J Cardiothorac Surg. 2020;4:e133–e41.

    Google Scholar 

  28. Scott A, Wade JBF, Wise AK, Shepherd RK, James NL, Stoddart PR. Force measurements at the tip of the cochlear implant during insertion. IEEE Trans Biomed Eng. 2014;61:1177–86.

    Article  Google Scholar 

  29. Wang P, Zhang S, Liu Z, Huang Y, Huang J, Huang X, et al. Smart laparoscopic grasper integrated with fiber Bragg grating based tactile sensor for real-time force feedback. J Biophotonics. 2022;15:e202100331.

    Article  PubMed  Google Scholar 

  30. Derossis AM, Fried GM, Abrahamowicz M, Sigman HH, Barkun JS, Meakins JL. Development of a model for training and evaluation of laparoscopic skills. Am J Surg. 1998;175:482–7.

    Article  CAS  PubMed  Google Scholar 

  31. Fried GM, Derossis AM, Bothwell J, Sigman HH. Comparison of laparoscopic performance in vivo with performance measured in a laparoscopic simulator. Surg Endosc. 1999;13:1077–81. discussion 82

    Article  CAS  PubMed  Google Scholar 

  32. Staveland SGHLE. Development of NASA-TLX (task load index): results of empirical and theoretical research. Adv Psychol. 1988;52:139–83.

    Article  Google Scholar 

  33. Liu W, Bretz F, Cortina-Borja M. Reference range: which statistical intervals to use? Stat Methods Med Res. 2021;30:523–34.

    Article  MathSciNet  PubMed  Google Scholar 

  34. Hopper AN, Jamison MH, Lewis WG. Learning curves in surgical practice. Postgrad Med J. 2007;83:777–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Szasz P, Louridas M, Harris KA, Aggarwal R, Grantcharov TP. Assessing technical competence in surgical trainees: a systematic review. Ann Surg. 2015;261:1046–55.

    Article  PubMed  Google Scholar 

  36. Rodrigues SP, Horeman T, Sam P, Dankelman J, van den Dobbelsteen JJ, Jansen FW. Influence of visual force feedback on tissue handling in minimally invasive surgery. Br J Surg. 2014;101:1766–73.

    Article  CAS  PubMed  Google Scholar 

  37. Wottawa CR, Genovese B, Nowroozi BN, Hart SD, Bisley JW, Grundfest WS, et al. Evaluating tactile feedback in robotic surgery for potential clinical application using an animal model. Surg Endosc. 2016;30:3198–209.

    Article  PubMed  Google Scholar 

  38. Sugiyama T, Lama S, Gan LS. Forces of tool-tissue interaction to assess surgical skill level. JAMA Surg. 2018;153:234–42.

    Article  PubMed  Google Scholar 

  39. Scott DJ, Dunnington GL. The new ACS/APDS skills curriculum: moving the learning curve out of the operating room. J Gastrointest Surg. 2008;12:213–21.

    Article  PubMed  Google Scholar 

  40. Smit D, Spruit E, Dankelman J, Tuijthof G, Hamming J, Horeman T. Improving training of laparoscopic tissue manipulation skills using various visual force feedback types. Surg Endosc. 2017;31:299–308.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to sincerely thank the gynecologists and other medical students who participated in this study.

Funding

The authors would like to thank the National Natural Science Foundation of China for funding (61975250).

Author information

Authors and Affiliations

Authors

Contributions

Yuxin Huang, and Jie Chen collected and analyzed the data. Qiongxiu Liao and Yuting Huang organized the participants to participate in the study. Xuemei Huang and Pingping Wang contributed to design the research and write the manuscript, Zhengyong Liu and Dongxian Peng were major contributors in revising and editing the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Zhengyong Liu or Dongxian Peng.

Ethics declarations

Ethics approval and consent to participate

This project was approved by the ethics Review Board of Zhujiang Hospital, Southern Medical University (2022-KY-081-01). Written informed consents were obtained from participants. All the methods and procedures carried out in this study were in accordance with relevant guidelines and regulation.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Huang, X., Wang, P., Chen, J. et al. An intelligent grasper to provide real-time force feedback to shorten the learning curve in laparoscopic training. BMC Med Educ 24, 161 (2024). https://doi.org/10.1186/s12909-024-05155-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12909-024-05155-1

Keywords